| Version 1.9 Build 803
|
|
Next: Example data and scripts
Up: VLA reduction
Previous: Basic Calibration Fundamentals
Listed below are the basic steps for typical continuum and spectral
line data reduction sessions which solve for and apply the standard
complex gain (G), the polarization leakage (D), and the relative
bandpass (B).
Since reduction of continuum, spectral-line, and polarimetry data in
AIPS++ differ essentially only in the addition or omission of
specific steps, and in the details of a few parameter settings, only
one pass through the basic reduction path is described here. In the
remainder of this document, annotated sequences of glish for both
continuum polarimetry and spectral-line reduction are included in
parallel. Some actions apply to only one type of reduction, but all
methods are instructive for either type of observation.
The basic reduction sequence is:
- Import the data. Data may be imported from either a VLA archive
tape using the vlafiller tool, or from a UVFITS disk file (e.g., as
written by AIPS) using the ms tool's ms.fitstoms tool constructor.
The result is a Measurement Set table (the AIPS++ internal data
format). The ms tool is used to obtain a summary of the dataset.
- Examine and edit the data using the flagger tool for basic data
editing, the autoflag tool for automatic data editing, and the
msplot tool for interactive editing.
- Set the initial calibration models. By default, all sources
have unit flux density point source models. Use the Imager.setjy function to set (point-source) flux densities for the absolute flux
calibrators.
- Obtain the complex gain (G) calibration for calibrators using
the calibrater tool. If planning to do instrumental polarization (D)
calibration, application of parallactic angle (P) corrections is
required at this point.
- Obtain the bandpass calibration (B) (spectral-line observations
only) for bandpass calibrator(s) using the calibrater tool. The G
calibration obtained above (and P, if necessary) is applied on-the-fly
to ensure that the B calibration is normalized.
- Transfer the flux density scale derived from the absolute flux
calibrator(s) using the calibrater tool. Only
the G calibration solutions obtained for the amplitude calibrators are
properly scaled in amplitude. The solutions for the other calibrators
are rescaled with the calibrater.fluxscale function by demanding that
the mean gain amplitude be the same for all calibrators.
- Apply the derived calibration to the source data.
- Obtain the instrumental polarization calibration (D)
(polarization observations only). Use a full-polarization model for a
calibrator (obtained by imaging1.1, and set with the Imager.setjy function).
All previous
calibration factors are applied before the D-terms or ``leakage
terms'' are determined.
- Apply all calibrations to the target source data. The
calibrater.correct function applies calibration to the
observed data and writes the result in the CORRECTED_DATA column in
the MeasurementSet.
- Image the target source. The imager tool is used to form
images from the calibrated data, deconvolve these images, and restore
the residual errors to the model image.
- As necessary, self-calibrate the target source data and reimage.
New and improved calibration may be obtained by repeating the initial
calibration step(s), this time using the model image obtained in the
first round of imaging. Re-image.
- Repeat the self-calibration as necessary. Self-calibration need
not be limited to only the G calibration factor; any solvable
calibration component (G, T, B, and/or D, in the present
example) may be self-calibrated.
Keep in mind that these examples are basic ones. More sophisticated
sequences are possible. For example, it may be desirable to obtain
bandpass calibration from an optimum calibrator, and to apply this
on-the-fly during full-out G calibration on a different calibrator.
This operation can be performed in the G and B calibrations steps
listed above for the bandpass calibrator only, then repeating the G
calibration step for the other calibrators with the B calibration
applied. In fact, any sequence of operations of this nature is
possible; the fundamental requirement is that calibration components
that significantly influence the one being solved for should be
applied prior to the solution.
Next: Example data and scripts
Up: VLA reduction
Previous: Basic Calibration Fundamentals
Contents
Please send questions or comments about AIPS++ to aips2-request@nrao.edu.
Copyright © 1995-2000 Associated Universities Inc.,
Washington, D.C.
Return to AIPS++ Home Page
2004-08-28