EMC test facility specification

EMC measurement standards

Conducted emission(EME)(EN 55 022) Conducted immunity(EMS)IEC 61000-4-6(NEN 11000-4-6)
Radiated emission(EME)pre-compliant to CISPR 16 Radiated immunity(EMS)IEC 61000-4-3

Technical specification

 

a.  Chamber size, 3.4m x 5.1m x 3.0 m (WxLxH)
b.  EM shielding following the EN 50147-1 test procedure:
· 10 KHz - 60dB min. (Magnetic field)
· 100 KHz - 80dB min. (Magnetic field)
· 1 MHz - 100 dB min. (Magnetic field)
· 10 KHz...30 MHz - 100 dB min. (Electric field)
· 30 MHz... 1 GHz - 100 dB min. (Plane wave)
· 1 GHz....18 GHz - 90 dB min. (Micro wave)
c.  full anechoic conducted immunity measurements in the frequency range 150 KHz - 230 MHZ, conform IEC 61000-4-6, RF power handling 50 Watt min.
d.   full anechoic radiated immunity measurements in the frequency range 30 MHz - 3 GHZ, conform IEC 61000-4-3 where applicable and capable of handling field strengths of 10 V/m min.
e.  full anechoic conducted emission measurements in the frequency range of 150 KHz - 230 MHz, conform EN 55 022.
f.  full anechoic radiated emission measurements in the frequency range of 30 MHz - 18 GHz, pre-compliance to CISPR 16.
g.  antenna to EUT distance 2 meter
h.  maximum EUT size (for field uniformity), 1.0 x 1.0 meter
i. Field uniformity TBD

EMC test equipment and accessories

EMC antennas

a. Schwarzbeck biconilog VULB-9163, 30-3000MHz
b. Schwarzbeck horn antenna BBHA-9120D, 1-18GHz

Radiated and conducted test equipment

a. Automatic RF Immunity test system mdl CRI-10CE, Comtest
b. RF power amplifiers:
· Prana, 10 KHz - 1.0 GHz, RF output power 50 Watt
· Milmega, 0.8 GHz - 2.5 GHz, RF output power 35 Watt
c. Dual RF power meter, Booton
d. Field calibration kit, Comtest
e. M3, 16 Amp. CDN network, Luthie
f. EM 101 Injection clamp, Luthie
g. M3 /EM-101, calibration kit, Luthie
h. SML-03 signal source, Rhode&Schwarz
i. FSP13 Spectrum analyzer, Rhode&Schwarz