Netherlands Institute for Radio Astronomy
| Conducted emission(EME)(EN 55 022) | Conducted immunity(EMS)IEC 61000-4-6(NEN 11000-4-6) |
| Radiated emission(EME)pre-compliant to CISPR 16 | Radiated immunity(EMS)IEC 61000-4-3 |
| a. | Chamber size, 3.4m x 5.1m x 3.0 m (WxLxH) |
| b. | EM shielding following the EN 50147-1 test procedure: · 10 KHz - 60dB min. (Magnetic field) · 100 KHz - 80dB min. (Magnetic field) · 1 MHz - 100 dB min. (Magnetic field) · 10 KHz...30 MHz - 100 dB min. (Electric field) · 30 MHz... 1 GHz - 100 dB min. (Plane wave) · 1 GHz....18 GHz - 90 dB min. (Micro wave) |
| c. | full anechoic conducted immunity measurements in the frequency range 150 KHz - 230 MHZ, conform IEC 61000-4-6, RF power handling 50 Watt min. |
| d. | full anechoic radiated immunity measurements in the frequency range 30 MHz - 3 GHZ, conform IEC 61000-4-3 where applicable and capable of handling field strengths of 10 V/m min. |
| e. | full anechoic conducted emission measurements in the frequency range of 150 KHz - 230 MHz, conform EN 55 022. |
| f. | full anechoic radiated emission measurements in the frequency range of 30 MHz - 18 GHz, pre-compliance to CISPR 16. |
| g. | antenna to EUT distance 2 meter |
| h. | maximum EUT size (for field uniformity), 1.0 x 1.0 meter |
| i. | Field uniformity TBD |
| a. | Schwarzbeck biconilog VULB-9163, 30-3000MHz |
| b. | Schwarzbeck horn antenna BBHA-9120D, 1-18GHz |
| a. | Automatic RF Immunity test system mdl CRI-10CE, Comtest |
| b. | RF power amplifiers: · Prana, 10 KHz - 1.0 GHz, RF output power 50 Watt · Milmega, 0.8 GHz - 2.5 GHz, RF output power 35 Watt |
| c. | Dual RF power meter, Booton |
| d. | Field calibration kit, Comtest |
| e. | M3, 16 Amp. CDN network, Luthie |
| f. | EM 101 Injection clamp, Luthie |
| g. | M3 /EM-101, calibration kit, Luthie |
| h. | SML-03 signal source, Rhode&Schwarz |
| i. | FSP13 Spectrum analyzer, Rhode&Schwarz |